Special Instructions for Using Electron and Photon Dosimetry Services
(46010C-48020S)
-
The NIST dosimetry calibration and test services for x rays, gamma rays,
beta particles, and electrons are performed in NIST's laboratories at
Gaithersburg, Maryland. Inquiries should be addressed to the appropriate
technical contacts. The
inquirer must provide the name and telephone number of an individual who can
answer technical questions that may arise in any inquiry, order, or shipment.
Upon receipt of an instrument, a report number is assigned. Calibrations are
generally performed in the sequence established by those numbers, except when
greater efficiency can be achieved by combining similar calibrations, or when
work for a calibration laboratory is given a higher priority. Arrangements
for calibration must be made in advance by letter, fax, e-mail or telephone,
so that the instrument or source to be calibrated will not be shipped to NIST
until the time of its scheduled calibration approaches. Inquiry should be made
as to scheduling and turn-around time.
Except for negligence by its personnel, NIST assumes no responsibility for
loss of or damage to the instruments or sources while in its possession. The
risk should be covered by insurance.
The report of calibration or test will carry a DG number (e.g., DG 9603/95).
Subsequent reference to that calibration or test should cite the DG number.
X-Ray and Gamma-Ray Measuring Instruments (46010C-46050S)
X-Ray Calibrations
X-ray measuring instruments are calibrated in terms of air kerma or exposure
by a substitution method in an x-ray beam at a point where the rate has been
determined by means of a free-air ionization chamber standard. In order to
provide instrument calibrations over a wide range of x-ray beam qualities,
many combinations of generating potential and filtration are available.
Tungsten (W) anode, x-ray beams with U.S. established beam qualities are
listed in Table VIII.5 as
lightly (L), moderately (M), and heavily (H) filtered beams. Two
beam qualities that do not fit into these categories are considered as
special (S) qualities. Cobalt-60 and cesium-137 gamma-ray beams are also
available. New W-anode, ISO x-ray beam qualities, listed in
Table VIII.6 are being
installed; check for availability. Molybdenum (Mo) and rhodium (Rh) anode
x-ray beam qualities, with application to mammography, are listed in
Table VIII.7. Beam
qualities are identified by beam codes given in the first column. The
calibration beam qualities requested should be appropriate to the instrument
submitted.
Gamma-Ray Calibrations
Gamma-ray measuring instruments are calibrated in terms of air kerma or
absorbed dose at points in the collimated cobalt-60 and cesium-137 gamma-ray
beams that have been standardized by means of graphite cavity chambers or a
graphite (or water) calorimeter. Rates at the time of calibration are computed
from the original beam standardization data and appropriate decay corrections.
Ionization chambers submitted for an air kerma calibration should have
sufficient wall thickness to provide electron equilibrium for the highest
energy selected. Ionization chambers submitted for an absorbed-dose
calibration must be suitable for calibration in a phantom.
Calibration Coefficient vs. Calibration Factor
The calibration quantity is defined as the conventional true value of the
quantity the instrument is intended to measure, divided by the instrument's
reading; this calibration ratio is termed a coefficient if it has
dimensions or a factor if it is dimensionless. An ionization chamber and
electrometer combination, with the electrometer scale in units of air kerma,
exposure, or absorbed dose, is calibrated by providing a dimensionless
calibration factor for the electrometer scale. An ionization chamber and
electrometer combination marked in electrical units is calibrated as follows:
(1) the chamber is calibrated in terms of air kerma or absorbed dose per
unit charge using an NIST electrometer; (2) the customer's electrometer is
checked for linearity and charge measurement accuracy; and (3) the
combination of chamber and electrometer is checked for consistency. An
ionization chamber submitted without an electrometer is calibrated in terms of
air kerma or absorbed dose per unit charge. Calibration can be based on
measurements for positive or negative polarizing potential, or on the mean of
measurements for both potentials, as requested. The ratio of ionization
currents for full and half polarizing potentials and the corresponding
ionization current will be stated in the calibration certificate, based on
pre-calibration measurements.
Calibration - Quality Chambers
Ionization chambers are tested, prior to calibration, for connection to the
atmosphere. Chambers found unsuitable for calibration will be returned with a
statement of the reason for rejection. A charge may be made for time incurred
on the tests. Each instrument submitted to NIST for dosimetry calibration or
test must be uniquely identified, usually by the manufacturer's name, model
number, and instrument serial number. When the serial number is lacking, an
alternative identifying mark should be provided. If none is found, NIST will
mark the piece with an identification number. If the apparatus submitted has
been calibrated previously by NIST, the serial number or identifying mark
should be given on the new order so that a continuing record of stability can
be maintained. A method of verifying instrument stability should be
established before shipment to NIST. Measurement should be made of the
instrument response both before and after shipment, using a long-lived
radioactive source and a highly reproducible measurement procedure. A
long-term record of instrument stability using a suitable constancy check
procedure is the most effective method for assuring the validity of the
instrument calibration.
Proficiency Tests
In an effort to meet the needs of calibration facilities which maintain
traceability with NIST for air kerma measuring instruments, NIST offers
proficiency tests upon request. NIST will conduct a proficiency test as
part of
an accreditation program or on an individual basis with a
calibration facility,
such as the biennial mammography
proficiency test
requirement of the FDA/MQSA. Inquiries should be addressed
to the appropriate technical contacts. The inquirer must provide the name and
email or telephone number of an individual who can answer technical questions
that may arise in any inquiry, order, or shipment. No NIST calibration service
ID number is associated with proficiency testing. The cost of proficiency
testing is the cost of x-ray and gamma-ray calibrations plus 15 percent.
Irradiations
Irradiation of passive dosimeters, for readout by the customer, is available
for most of the beam qualities listed in
Table VIII.5. These
irradiations are generally in terms of air kerma; for passive dosimeters
suitable for insertion in a phantom, irradiation in terms of absorbed dose can
be provided by in-phantom irradiation using cobalt-60 gamma rays.
Calibrations of x-ray and gamma-ray measuring instruments and of passive
dosimeters, described above, have a relative expanded uncertainty of
1 %, based on a coverage factor of k = 2 to define an interval
having a level of confidence of approximately 95 %.
Gamma-Ray Sources, Beta-Particle Sources, and Measuring Instruments
(47010C-47040S)
Sources submitted to NIST for dosimetry calibration are subject to the
following conditions:
- Preparation: Sources submitted for calibration must be sealed so that
there can be no escape of any radioactive material, including any gaseous
decay products. The sources, shielding, and packaging must be free of
contamination. Contaminated or leaking sources cannot be measured and may
cause considerable loss of time and damage to laboratory facilities. Sources
must have been sealed for a sufficient time to be substantially in radioactive
equilibrium with their decay products when these contribute to the emitted
radiation.
- Packaging for shipment: Packages must be in compliance with the
regulations of the Department of Transportation as specified in
DOT 49CFR173.401-173.476. Radionuclides must be packaged as Limited
Quantities (DOT 49CFR173.421-173.422) or in Type A packages
(DOT 49CFR173.412 and 173.433). Type A packages must bear the
appropriate radioactive-hazard labels (DOT 49CFR172.403). If the source
is considered by the shipper to be in DOT Special Form, a Special Form
certificate must be furnished to NIST in strict compliance with
DOT 49CFR173.476. Copies of the codes are available from the Government
Printing Office, Washington, DC 20402.
All shipments to NIST of gamma-ray and beta-particle sources should be in
reusable containers. A drawing showing the source container and a description
of the method of source removal should be provided before the shipment is
received at NIST.
If the nature of the shipment requires a Type B container subject to an
NRC quality assurance program, documentation should be supplied to NIST
certifying that the use of the container by NIST is part of the program of
the shipper.
- Possession of licensed materials: In submitting a source for calibration,
it is necessary for the submitter to certify that he is duly authorized to
possess the source under license by the applicable authority. In the case of
individuals residing in a State that has entered into agreement with the
Nuclear Regulatory Commission, State regulations are applicable to all
sources. In the case of other individuals, NRC regulations are applicable.
This certification may be by letter, by a suitable statement on the purchase
order covering the calibration fee, or by a clear copy of the submitter's
Possession License for the source.
Calibration in terms of air-kerma strength (air-kerma rate in free space times
the square of the distance of the calibration point from the source center
along the perpendicular bisector) is provided for gamma-ray sources of
cobalt-60, cesium-137, iridium-192, and iodine-125. Calibration in terms of
absorbed-dose rate is provided for suitable encapsulated beta-particle
sources; the dose rate to a low-atomic-number material (graphite or plastic)
is determined by measurement with an extrapolation chamber. The beta-particle
sources may be either small-area sources such as ophthalmic applicators, or
large-area plaques, and will be calibrated for absorbed dose rate to water
either at the source surface or at a specified distance.
Ionization chambers to be calibrated with beta-particle sources must be
parallel-plate chambers with thin walls. They can be calibrated with the
radionuclides 90Sr + 90Y, or 204Tl,
or 147Pm.
Measurement services in this series have uncertainties listed in
Table VIII.8 and
Table VIII.9.
Dosimetry of High-Energy Electron Beams (48010M-48020S)
-
Alanine dosimeters are provided twice a year to users requesting assistance
with absorbed-dose measurements in high-energy electron beams. The user
irradiates all but one of the three furnished dosimeters to between
50 Gy and 80 Gy (5000 rad and 8000 rad) to water at
electron energies between 5 MeV and 50 MeV, employing the irradiation
geometry (field size, phantom, position of dosimeter in phantom) given in the
"Protocol for Dosimetry of High-Energy Electrons," Phys. Med.
Biol. 11, 505 (1966).
After irradiation, the dosimeters are returned to NIST for evaluation terms of
absorbed dose in the phantom, using appropriate methods. These dose
interpretations ignore certain corrections for the effects of spectral
perturbations, and so represent a measurement quality assurance service rather
than a calibration service. The measurement procedure with alanine dosimeters
is undergoing refinement; for the current procedure, the estimate of the
relative expanded uncertainty is 2.3 %, based on a coverage factor of
k = 2 to define an interval having a level of confidence of
approximately 95 %.
Inquiries or comments:
s.seltzer@nist.gov
Online: June 1996 - Last update: March 2005
|