Technical Highlights
- International Comparison of Mammographic X-ray Exposure Standards.
In June of 1998 NIST completed a direct comparison of primary standards for
mammographic x-rays at the National Physical Laboratory (NPL) in Teddington,
UK. The Ritz chamber, a NIST free-air ionization primary standard used between
20 and 100 kV was compared to the NPL primary standard used for
calibrations of mammography x-ray qualities. Good agreement, between 0.4 %
and 0.6 %, was found between the two primary standards. In 1997 an
indirect comparison with the German primary standard for mammography showed
similar results to the recent NPL-NIST comparison. This indirect comparison was
conducted through the use of NIST reference class ionization chambers, which
were calibrated at NIST and at the German standards laboratory. The results
from the calibrations at both national laboratories agreed to within 0.5 %
for the six different mammographic beam qualities used in this comparison. The
NIST primary standard for mammography x-rays, the Attix chamber, was also
directly compared with another NIST national standard for low energy x-rays,
the Lamperti chamber. The results from the comparison of the two primary
standard chambers agreed to within 0.4 %, for the four molybdenum entrance
x-ray beam qualities. (C.M. O'Brien and P.J. Lamperti)
- Implementation of ISO Bremsstrahlung and NIST Techniques. NIST has
continued the efforts to implement the photon techniques specified by the
International Organization for Standardization (ISO). In addition to the
implementation of the 41 new ISO beam qualities, NIST has responded to the
requests of the medical community through the development of several new
diagnostic x-ray beam qualities. The development project of the diagnostic
M80 and M120 beam qualities was identified by CIRMS as a high-priority need
and will soon be successfully completed. Calibrations to the ISO beam
qualities and the diagnostic beam qualities will be available upon the
completion of an evaluation of the correction factors for the primary
standards used to measure these new beam qualities. A spectral determination
of all new beam qualities will be used to verify the half-value layers of the
newly established beam parameters. (C.M. O'Brien, P.J. Lamperti, and
S.M. Seltzer)
- International Comparisons of Low-Energy X-ray Primary Standards.
Low-energy x-ray comparisons of two NIST primary standards were conducted at
both the National Physical Laboratory (NPL) in Teddington, UK, and at the
Bureau International des Poids et Mesures (BIPM) in Sevres, France, in the
summer of 1998. The NPL-NIST 1998 low-energy portion of the comparison
involved a series of air-kerma measurements with both the NIST Lamperti and
Ritz free-air chambers and comparisons with results from the NPL
1 cm3 standard chamber. The NPL x-ray beam qualities used for
the comparison were produced between 10 kV and 50 kV. Agreement was
found to be within 0.5 %. The diagnostic-energy portion of the
NPL-NIST 1998 comparison involved a series of air-kerma measurements with
the NIST Ritz chamber and the new NPL high-energy standard chamber at beam
qualities produced at 50 kV and 80 kV. Preliminary results indicate
agreement to within 0.6 %. The BIPM-NIST 1998 comparison included a repeat
of air-kerma measurements made in 1966 with the NIST Lamperti chamber. The goal
was to replicate the measurement conditions of the 1966 comparison without
compromising the experiment. Although charge-measurement techniques have
changed since 1966, the beam-quality parameters, chamber apertures and
alignment techniques have remained unchanged at BIPM. The five BIPM beam
qualities for the Lamperti chamber comparison were produced by a tungsten
anode at 10, 25, 30, and 50 kVp. The NIST Lamperti chamber reproduced the
results of 32 years earlier to within 0.1 %, and agreement with the
BIPM standard in this range was found to be within 0.6 %. Comparisons of
results from the NIST Ritz chamber and the BIPM standard were made at six BIPM
tungsten beam qualities produced at 25, 30, 50, 80 and 100 kVp.
Preliminary results indicate agreement that varies from 0.4 % to
1.0 %. (C.M. O'Brien and P.J. Lamperti)
- Photon and Charged-Particle Data Center. The Data Center compiles,
evaluates, and disseminates data on the interaction of ionizing radiation with
matter. The data on photons and charged particles, with energies above about
1 keV, include fundamental information on interaction cross sections as
well as transport data pertaining to the penetration of radiation through bulk
material. Databases are developed and maintained on attenuation coefficients
for x rays and gamma rays, including cross sections for Compton and
Rayleigh scattering, atomic photoeffect, and electron-positron pair
production, as well as on energy-transfer, energy-absorption and related
coefficients relevant to radiation dosimetry. Work on charged-particle cross
sections and of radiation transport data has entailed significant effort on
the evaluation of the stopping powers and ranges of electrons, positrons,
protons, and alpha particles, the elastic scattering of electrons and
positrons, and the cross section for the production of bremsstrahlung by
electrons. The quality of the work of the Data Center is reflected in the many
requests for our data from other laboratories and in the use of our data in
engineering and scientific compendia, books and review articles, and in the
reports and protocols of national and international standards organizations.
The compilations of the Data Center rely heavily on the synthesis of available
theory to extend the data and provide for comprehensive coverage over broad
ranges of energy and materials. Thus we have long been involved in complex
computational analyses and in the development of highly sophisticated
transport-theoretic methods. Our Monte Carlo transport calculations also are
incorporated into some of the most widely used general-purpose radiation
transport codes. With the help of the NIST Physics Laboratory's Office of
Electronic Commerce of Scientific and Engineering Data, a number of the
Center's databases can be accessed on the worldwide web. (S.M. Seltzer,
J.H. Hubbell, and M.J. Berger).
- Theoretical Dosimetry for Brachytherapy Sources. Brachytherapy
sources, encapsulated radioactive material emitting photon and/or beta
particles, are used in a variety of therapeutic applications. Such sources
include ophthalmic applicators for the treatment of ocular lesions, low-energy
photon "seed" implants for the treatment of prostate cancer, and
photon- and beta-emitting intravascular sources of various design intended for
the irradiation of artery walls to prevent re-closing (restenosis) after
balloon angioplasty. Numerous calculations have been done for a wide variety
of source designs to determine the spatial distribution of absorbed dose per
unit activity in the source to aid in source optimization, to provide basic
dosimetric data, and to determine correction factors for measured results.
These calculations are done with state-of-the-art Monte Carlo codes, such as
ITS, MCNP4b, and EGS4, and with more rapid point-kernel-based methods.
(S.M. Seltzer, C.G. Soares, and F. Mourtada)
- 103Pd Brachytherapy Source Characterization. There has
recently been a significant increase in the demand for low-energy photon
brachytherapy sources used in the treatment of prostate cancer. We are working
with source manufacturers and the medical physics community in order to
develop a new standard for the dosimetry of brachytherapy seeds containing the
radionuclide 103Pd. Measurement of the air kerma strength of
103Pd seeds is accomplished using the Wide-Angle-Free-Air Chamber
(WAFAC). Ionization current measurements using well-ion chambers and
Kα photon fluence measurements are also carried out on each
seed. Various ratios of these quantities are computed and evaluated for
suitability as calibration factors for source manufacturer quality control
measurements and source user treatment planning. (M.G. Mitch,
P.J. Lamperti, B.E. Zimmerman, F.J. Schima, C.G. Soares,
S.M. Seltzer, and B.M. Coursey)
- Brachytherapy Source Dosimetry Using the Imaging Plate. The
continuing development of new source geometries with lower required activities
to achieve desired doses has prompted an investigation of the possible
utilization of Fuji plate phosphoimaging to characterize weak brachytherapy
sources. The Imaging Plate contains a photo-stimulable phosphor which "stores"
energy from radiation sources. Following exposure, the plate is placed into a
reader, which scans the plate surface with a laser, releasing photo-stimulated
luminescence, which is converted into a two-dimensional digital image. A
preliminary depth-dose study using a 32P intravascular
brachytherapy wire source in an A150 plastic photon showed the ability of the
imaging plate to successfully measure dose rates in the µGy/s range, below that
accessible with radiochromic film. (M.G. Mitch and C.G. Soares)
- Intravascular Brachytherapy Source Dosimetry. The use of
beta-particle emitting brachytherapy sources for the prevention of restenosis
(re-closing) of coronary blood vessels after angioplasty continues to be
actively explored. NIST has taken an early and leading role in the calibration
of the sources used for this therapy, employing the NIST extrapolation chamber
equipped with a 1 mm diameter collecting electrode to measure dose rate
at a depth of 2 mm in water-equivalent plastic. These measurements are
confirmed using radiochromic dye film, which is also used to characterize
sources in the cylindrical geometry for transaxial uniformity. In addition,
irradiation of planar sheets of film at various depths in water-equivalent
plastic were used to construct data sets which can be used to predict the dose
rate at arbitrary locations around the sources using a modified form of the
AAPM Task Group 43 Protocol. A publication describing this work was
published in the journal Medical Physics. The equipment used for these studies
was augmented with the addition of an automated micro-scintillator detection
system and various well-ionization chambers. Use of all this equipment was
centralized in a newly refurbished laboratory. Collaborations were continued
between NIST and NeoCardia for dosimetry of a 32P wire, with
Washington Hospital Center for dosimetry of various sources, and with Radiance
Medical Systems for dosimetry of a radioactive balloon source. Collaborations
were also begun with Best Industries for 188W/Re wire sources.
Funding was received from the National Advanced Manufacturing Testbed (NAMT)
Program at NIST for this work, with the emphasis on the automation of the
source handling for the measurements. An international workshop on
brachytherapy source dosimetry was organized and hosted by NIST in April 1998.
(C.G. Soares and M.G. Mitch, with F.A. Mourtada,
T. Wheatley, and R. Densock, Div. 823)
- International Gamma Ray High-Dose Comparison. The national
metrology institutes of the United States (NIST), France (BIPM), Great Britain
(NPL), Germany (PTB), Italy (INMRI-ENEA), and the International Atomic Energy
Agency, have undertaken a high-dose comparison. The BIPM coordinated the
comparison and will analyze and publish the results. The protocol was created
through consultation with the two issuing laboratories, NIST and NPL. The
dosimeters have been processed by the irradiating laboratories and are
currently being measured by the issuing laboratories. (M.F. Desrosiers
and V. Nagy)
- Internet-Based Calibration Services for the Radiation-Processing
Industry. In cooperation with industry, academia and other government
agencies, an Internet-based system will be built for fast remote calibration
of high-dose radiation sources against the U.S. national standard
gamma-radiation source. The new project will be facilitated by the National
Advanced Manufacturing Testbed (NAMT) based in NIST's Manufacturing Engineering
Laboratory. The new service will deliver immediate calibration results to the
industrial customer on-demand at a lower cost. All prerequisites are now
present for automating NIST's calibration services on the basis of modern
technologies and commercially available products. Once built, the
Internet-based transfer calibration process will be the most modern
calibration service available and offer many options for expansion into other
areas of dosimetry and metrology in general. The new service will entirely
redesign the source calibration process to make it faster, less laborious,
much cheaper, and, consequently, much more accessible for present and
potential customers. (M.F. Desrosiers, J.M. Puhl, S.M. Seltzer,
T. Vorburger, and T.B. Renegar)
- Validation of the EPR Method for Tooth Enamel Dosimetry. The main
objective of radiation protection is the development of systems of protection
and decision making that help minimize unfavorable consequences of
occupational and non-occupational human exposures. To meet this objective,
knowledge is required on dose-effect relationships for radiation-induced,
stochastic and deterministic effects. Therefore, the acquisition of dosimetric
data from populations with chronic exposure is of special interest. Several
new sources (Chernobyl, Techa River) for these studies have arisen and are
presently underway. Electron Paramagnetic Resonance (EPR) is the only physical
method available to retrospective biological dosimetry studies. Validation of
the method and rigorous analysis of critical steps in the method have not been
demonstrated in the scientific literature. Validation is essential before
these data can be used reliably in epidemiological studies from which
recommendations are made for occupational exposures. The objective of this
work is to validate the EPR-tooth dose assessment method and produce a
standardized procedure with well-defined uncertainties. (M.F. Desrosiers,
O.F. Sleptchonok, V. Nagy)
- Alanine-EPR Film Dosimeter. Prototypes of a new polymer-based film
dosimeter containing alanine have been manufactured and tested. This
development program, under a CRADA with the W.R. Grace Co., is testing a
variety of formulations for the film dosimeter. These tests were followed by
comparative measurements on films of the same polymer/alanine ratio
manufactured by different methods. The selected film formulation was
mass-produced and is undergoing extensive tests of its radiation response
characteristics. Using NIST accelerators and with the cooperation of
Risø National Laboratory and the Hungarian Academy of Sciences, a study
of the energy response of the films confirmed previously published data on
alanine. (M.F. Desrosiers, V. Nagy, W.L. McLaughlin,
C.E. Dick, and J.M. Puhl)
- Production and Characterization of Radioendofullerenes. An
investigation of novel, fullerene-based, delivery methods for radionuclides in
medical/industrial imaging and tracing applications is in progress.
Radioendofullerenes have radionuclides trapped inside their all-carbon,
cage-like structures, isolating chemically reactive isotopes from the
environment. The Kratschmer-Huffman method of fullerene synthesis was carried
out in a Fullerene Production Chamber (FPC) using high-current arc burning of
graphite rods doped with the isotope to be encapsulated. High-pressure liquid
chromatography (HPLC) and thin layer chromatography (TLC) prior to
identification of various, purified, fullerene species by optical absorption
spectroscopy and mass spectrometry. Current studies have focused on the
encapsulation of iodine, as it has several medically useful isotopes.
Following development and execution of an encapsulation protocol for
"cold"" 127I, the radionuclide 126I was
produced by the reaction 127I(γ,n) 126I upon
irradiation of the sample with bremsstrahlung from our MIRF (Medical Industrial
Radiation Facility) linac. The radiolabel was identified by MultiPhoton
Detection (MPD) and Fuji plate phosphoimaging. Direct encapsulation of the
radionuclide 125I has recently been accomplished, and
characterization of this new radioendofullerene species is currently in
progress. (M.G. Mitch, L.R. Karam, B.M. Coursey, and
K.L. Rowley)
- Neutron Interferometry and Optics Facility (NIOF). During the past
year a diverse selection of experiments in fundamental and applied physics
have been successfully carried out at the NIOF. These experiments have so far
resulted in the writing of six articles for publication and have partially
satisfied requirements for two separate Ph.D. dissertations. This work has
involved collaborations with the U. Missouri-Columbia, the Hahn-Meitner
Institute, the U. Innsbruck, the Exxon Research and Engineering
Corporation, and the NIST Polymers Division.
One of these experiments was aimed at searching for the recent predictions of
quantum entanglement of the nuclear states in a mixture of fluids. The
existence of such entanglement would suggest that the neutron refractive
index, n, of a mixture cannot be calculated from the knowledge of
refractive indices for the constituent elements alone. The experiment measured
n for various mixtures of H2O and D2O. Some
theorists have predicted a 5 % to 10 % deviation from the traditional theory due
to quantum entanglement of H and D at room temperature. However, the NIOF
experimental data agreed with the traditional theory to within the statistical
precision of the data, which was of order 0.4 %.
The first, successful, neutron interferometric measurement of the mass density
of a thin polymer film (thickness <1 µm) was made at the NIOF in
collaboration with the NIST Polymers Division. This new technique to measure
thin film density is self-calibrating and does not require complex,
mathematical modeling of the makeup or surface profile of the film.
The capabilities of the NIOF have been augmented by the addition of a
transmission neutron polarizer and a RF gradient spin flipper. This polarizer
provides the NIOF with neutron polarization in excess of 98 %. This new
capability was exploited in an experiment that directly demonstrated the
4π spin rotation symmetry of the neutron wave function under space rotation.
This experiment was unique in that the neutron guide field was gently rotated
by 180°, allowing the neutron spin direction to adiabatically follow the field.
The measured phase shift is strictly due to the rotation of the spin direction
since it is independent of both the strength of the magnetic field and the
magnitude of the neutron magnetic moment.
A new, neutron diffraction, radiographic and tomographic, imaging setup has
been constructed, installed, and successfully tested. Along with this new
addition to the NIOF there exists a 2-dimensional, high resolution, CCD,
neutron detector. In addition to radiography, many other types of diffraction
imaging experiments were performed with this setup. Researchers from Exxon
exploited this facility to study hydrogen distribution in operating fuel
cells. Researchers from the U. Innsbruck, Austria, studied the diffraction
of neutrons (λ=0.235 nm) from macroscopic objects ~0.1 mm in
size. Finally, the Neutron Phase Contrast Imaging technique was demonstrated
by a high school student, resulting in his recognition as a semi-finalist in
the 57th Westinghouse Science Talent Search. (M. Arif,
D. Jacobson, A. Ioffe, P. Huffman, T. Gentile, and
A. Thompson)
- Polarized 3He for Neutron Spin Filters and MRI
Applications. The primary focus of the polarized 3He program is
the development and application of neutron polarizers for both condensed
matter and fundamental physics. As a spin-off of this technology, the program
also includes collaboration with medical researchers in polarized gas magnetic
resonance imaging (MRI). A unique feature of the 3He effort is the
concurrent development of the spin-exchange and metastability-exchange optical
pumping methods. These two methods have different strengths and weaknesses,
and pursuit of both allows for a versatile approach to the needs of different
applications.
Neutron polarizer tests have been performed with a compact neutron polarizer
test facility located on the new NG6 monochromatic beamline. This apparatus
has a laser system for on-line, spin exchange, optical pumping of cells,
whereas cells polarized by the metastable method are presently transported
from another lab to the test facility. For the metastable method, polarized
gas is compressed using a modified diaphragm pump.
The first tests of polarizing neutrons have been performed on this apparatus.
A spin-exchange cell produced 88 % neutron polarization of the NG6M beam
with 8.1 % transmission. For a metastable-exchange cell, the results were
67 % neutron polarization with 18.5 % transmission. For both methods,
the 3He polarization was 45 %; the difference in the results
for neutron polarization and transmission are related to the
"thickness" of gas.
In collaboration with C. Glinka, J. Barker, B. Hammouda, and
J. Lynn of the NIST Center for Neutron Research, we have performed the
first application at a U.S. neutron facility of polarized 3He to
condensed matter research. The purpose of this test was to evaluate the use of
a 3He-based polarization analyzer in a small angle neutron
scattering (SANS) apparatus. Conventional neutron polarizers are not practical
for this goal because of the relatively high divergence of the scattered beam.
The experiment successfully demonstrated the separation of coherent from
incoherent scattering using polarization analysis, which had not previously
been performed in SANS experiments.
We have performed collaborative experiments in polarized gas MRI with the
U. Pennsylvania and the U. Virginia. Until our recent work, there
had been no application of the metastable-exchange method to polarized gas
MRI. We have obtained four lung images, three of patients with lung disease,
by using the compression apparatus to polarize 3He gas at NIST, and
then transporting the gas to U. Pennsylvania and U. Virginia.
(T. Gentile, A. Thompson, G. Jones)
- Asymmetries of the Weak Interaction and the Neutron Lifetime.
During FY98, two different neutron lifetime experiments, one a NIST-led, cold
neutron beam experiment, the other a Harvard-led, ultra cold neutron (UCN)
experiment, were conducted on our polychromatic neutron beam at the NIST Cold
Neutron Research Facility. On our adjacent monochromatic neutron beam, work
was begun to calibrate, using a lithium calorimeter, the efficiency of the
neutron fluence monitor belonging to the neutron lifetime experiment.
The NIST lifetime experiment utilizes a continuous, cold, neutron beam, a
Penning trap to catch decay protons, and a thin foil, neutron fluence monitor
to measure the lifetime. During FY98, 11 series of runs were completed. The
statistical uncertainty achieved in those runs was an impressive
1.9 seconds. Nevertheless, our greatest success came in a better
understanding of the systematic errors that have influenced results in the
past. Major problems affecting stability and reliability of voltages in the
Penning trap were identified and eliminated. Proton backscattering from the
solid state proton detector is another major concern in this experiment. A
correction is possible only when the dead layer of the detector has been
characterized.
The UCN lifetime experiment exploits three-dimensional, magnetic trapping of
UCN produced by inelastic scattering of cold neutrons in a reservoir of
superfluid 4He. As the trapped neutrons decay by beta emission, the
energetic electrons generate scintillations in the liquid He that can be
detected with nearly 100 % efficiency. A measurement of the scintillation rate
as a function of time is used to determine the neutron beta-decay lifetime.
The apparatus was moved onto the beamline late in 1997. Since then two
surprisingly strong sources of background have been studied: a constant
background rate and a time dependent rate coming from a trapped species other
than neutrons. Work is presently underway to remove these signals. The
constant background will be reduced by using additional neutron shielding,
muon veto counting, and coincidence between two photomultiplier tubes.
Indications are that the coincidence method of running will also reduce the
trapped signal to acceptable levels. (M.S. Dewey, J. Adams,
M. Arif, T. Gentile, D. Gilliam, P. Huffman,
D. Jacobson, G. Jones, J. Nico, A. Thompson, and
F. Wietfeldt).
- Neutron Dosimetry for Reactor Safety Assessment. The Neutron
Interactions and Dosimetry Group provides metrology services and research
studies on neutron metrology to the USNRC Office of Regulatory Research in
support of their program in nuclear reactor safety assessment. We are
presently completing a multinational, round-robin intercomparison of
fissionable, neutron fluence monitors of a design currently employed by the
nuclear industry for measuring the RPV fast neutron fluence. In particular,
the fission reaction in 237Np fissionable dosimeters is especially
useful because it monitors a significantly larger portion of the neutron
damage spectrum than do other dosimetry materials. However, several
complicating factors (e.g., the presence of a substantial 233Pa
equilibrium activity) contribute to considerable differences observed in the
results for the fast-neutron fluence obtained from neptunium dosimeters, as
compared to other dosimeters. From these tests we will be able to specify
details for the accurate use of 237Np as a neutron damage monitor.
As a complementary study, a round robin test of 93Nb dosimeters is
planned for 1999. (J. Adams and D. Gilliam)
- Neutron Cross Section Standards. NIST has played an important role
in the improvement of the neutron cross section standards through both
evaluation and experimental work. We are leading an effort that will result in
a new international evaluation of the neutron cross section standards. This
has involved motivating and coordinating new standards measurements, examining
the standards database, and pursuing the extension of the standards over a
larger energy range. This work has taken place through participation in the
U.S. Cross Section Evaluation Working Group and two international committees,
the International Nuclear Data Committee and the Nuclear Energy Agency Nuclear
Science Committee. An objective is to complete the evaluation in time for the
major international cross section evaluation projects to use the improved
standards in forming new versions of their libraries.
NIST has continued to maintain a limited experimental role in the measurements
of the standards. This role has led to a new, NIST-LANL-Ohio University
collaborative measurement of the H(n,n) angular distribution at Ohio
University at 10 MeV neutron energy. Preliminary results indicate
differences with the most recent U.S. evaluation of this angular distribution.
This work was initiated as a result of concern about that evaluation expressed
by European standards groups. The H(n,n) angular distribution is one of the
most widely used cross section standards. Also, plans are being made for a new
measurement of the 10B total cross section at our new,
monochromatic, neutron beam facility on NG6. This work will support efforts to
improve the 10B(n,αγ) standard at low neutron energies.
(A. Carlson)
- Dissemination of National Standards of Radioactivity. The
Radioactivity Group disseminated the national standards of radioactivity as
follows: 1) over 500 radioactivity Standard Reference Materials (SRMs)
were provided to customers; 2) over 200 comparative measurements and
reports of traceability were provided to federal regulatory agencies,
radiopharmaceutical manufacturers, commercial suppliers of calibration sources
and services, and the nuclear-power industry. Industrial steering committees
guided the work of four research associates in cooperative testing programs;
and 3) over 60 calibrations of customer sources were provided.
(L.L. Lucas, J.C. Cessna, and L.R. Karam)
- Glow-Discharge Resonance Ionization Mass Spectrometry (RIMS). Work
continued on the development of a glow-discharge initiated, mass spectrometer
system that would permit the direct, compositional analysis of soils and
sediments for radioactive and non-radioactive elements. The mass spectrometer
was upgraded with improvements in the vacuum systems, improvements in the ion
collection systems, renovation of the Faraday Cup insertion and control
system, purchase and installation of several needed electronic components, and
development of software programs for the automatic manipulation of various
physical parameters. Tests were performed with the glow-discharge source aimed
at maximizing the neutrals-to-ions ratio for cesium. The response of the
output to changes in the glow-discharge current, the internal pressure of the
argon gas in the source, and source construction and configuration were
examined and compared with theoretical expectations. Much work was performed
to obtain the optimal setup for resonance ionization with the TiSapph coupled
with an Ar ion laser for the ionization stage. The purchase of appropriate
optical parts enabled optimal focusing and alignment of the lasers. After
these system improvements were made, the effort was turned to developing a RIS
ion beam. Cesium was chosen as a first source element partly because of its
low ionization potential, which could lead to less demanding proof of
principle and, hence, the exploration of optimal conditions on critical
parameters. Initially, a RIS response using a thermal ionization source was
obtained using a double filament that allowed vaporization of the source
material at low enough temperatures to remove mainly the ion component and, at
the same time, allow large amounts of neutrals to be produced. Subsequently, a
RIMS beam was produced with the glow-discharge source. Work is now underway to
optimize the sensitivity and selectivity of this process. (L.R. Karam,
J.M.R. Hutchinson, and L. Pibida)
- Environmental Radioactivity SRMs for Radiochemical Metrology. Over
the past decades, on the order of 1015 Bq of waste have been
stored in the oceans by the nuclear countries. Potential contamination by
leaking nuclear waste, including the 17 sunken Russian nuclear submarines in
the Arctic Ocean, has caused worldwide concern for the exosphere and
commercial food chain. The determination of mBq/g levels of radioactivity in
ocean sediments is a difficult task. The NIST Ocean Sediment Standard
Reference Material (SRM 4357) now serves as the basis for all measurement
assurance programs, methods verification procedures, and data comparisons.
This SRM was developed using a composite of 0.5 % contaminated Irish Sea
sediment and 99.5 % Chesapeake Bay sediment by weight. Ten radionuclides,
including 40K, 90Sr, 137Cs, 226Ra,
228Ra, 228Th, 230Th, 232Th,
238Pu, and 239-240Pu, were certified through a
comparison study among 20 international laboratories. The mean values were
reported for an additional 11 uncertified radionuclides: 129I,
155Eu, 210Po, 210Pb, 212Pb,
214Bi, 234U, 235U, 238U,
237Np, and 241Am. Interestingly, the measurement results
for the primordial radionuclides were distributed normally, while those for
the anthropogenic radionuclides were best described by Weibull distributions.
SRM 4357 will be followed by additional low-level radioactivity natural
matrix SRMs. The Ashed Bone matrix was certified in 1998 and will be followed
by Rocky Flats soil II (1999) and the Japan, White, and Irish Seas
composite Shellfish. (Z.C. Lin and K.G.W. Inn)
- Radionuclide Speciation in Soils and Sediments. Potentially,
billions of dollars of remediation costs could be saved and social concerns
could be addressed with reliable and interpretable information on the chemical
forms (speciation) of the nuclides and their distribution in the environment.
The Ionizing Radiation Division will be developing a new series of
environmental SRMs for radionuclide speciation. As a first step towards this
goal, the Division, in collaboration with Florida State University, is
developing a phase-selective, standard, extraction protocol. The future,
environmental, radionuclide SRMs will be characterized for radio-plutonium,
uranium, thorium, strontium, and cesium species and will be used by those
making decisions for risk assessment, selection of cost-effective, mitigation
strategies and technologies, and developing and executing long-term
monitoring. The effects of critical variables (time, temperature and reagent
concentration) of the leaching protocol are being evaluated for selective
dissolution of the exchangeable, carbonate, reducible metal oxide, organic,
and acid leachable phases. Once the protocol is documented and pilot tested, a
network of collaborations will be developed for the future round-robin testing
of the protocol and interlaboratory comparisons for the certification of the
future SRMs. (K.G.W. Inn).
- 239Pu Intercomparison of ICP-MS, TIMS and FTA at µBq
Levels.. The goal of this project was to evaluate the state-of-the-art
accuracy (precision, and minimum detection amount) for 239Pu in
artificial urine measurements by inductively-coupled plasma, thermal
ionization, mass spectrometry and fission track analysis in the range of 18 to
278 µBq/sample (approximately 3 million atoms of plutonium per
200 g sample). The plan for diluting NIST 239Pu SRM from
kBq/kg to µBq/kg in artificial urine was developed by NIST and the Yankee
Atomic Environmental Lab. With careful selection of reagents, labware, and
high precision metrological techniques, the test materials were made, their
concentrations confirmed, and samples sent to participating laboratories.
Analysis of the data indicated that ICP-MS currently is capable of making
quantitative measurements of 239Pu from urine down to the
18 µBq/sample level within an uncertainty of 20 %. Some potential,
technical difficulties by some of the participating laboratories include:
1) data transcription; 2) sample tracking; 3) estimation of
limits of detection; 4) statistical control over analytical processes;
and 5) control of analytical background. A second study is being planned
to include interferences in order to assess the technologies under more
realistic conditions. (K.G.W. Inn)
- Experimental Determination of Dose Calibrator Settings for
188Re. Rhenium-188 is proving to be one of the most versatile
of the radionuclides currently being investigated for use in nuclear medicine.
Among the applications being studied for 188Re are bone pain
palliation therapy, radioimmunotherapy, radiation synovectomy, and, more
recently, intravascular brachytherapy (IVBT) for the prevention of coronary
restenosis following balloon angioplasty. Radioassays are normally performed
in the clinic using re-entrant well ionization chambers ("dose
calibrators") using a dial setting, or "calibration factor,"
recommended by the manufacturer. Accurate measurements of radionuclides in the
clinic or radiopharmacy therefore demand that the correct calibration factor be
applied. The dose calibrator settings for the medical radionuclide
188Re (as 188ReO4-) were
experimentally determined using solution sources prepared and calibrated by the
Radioactivity Group. These new settings result in activity readings 28 %
to 30 % lower than those obtained using the manufacturer's recommended
setting. This discrepancy most likely results from an underestimation of the
total radiation yield from 188Re decay by the manufacturer when
calculating the dose calibrator response to derive the dial setting. This study
emphasizes the need for experimental determinations of dose calibrator settings
in the geometry in which the measurements will be performed.
(B.E. Zimmerman and J.T. Cessna)
- A New Technique for Measuring Contained Activity in Balloon Catheters
for Use in Intravascular Brachytherapy: Radioassays and Film Dosimetry of
133Xe. Accurate dosimetry of liquid- or gas-filled balloon
catheters for use in intravascular brachytherapy requires accurate knowledge
of the amount of contained activity. However, assumptions made in estimating
this quantity often lead to serious errors. The Radioactivity Group has
developed a technique to make direct measurements of contained activity in
balloon catheters using a commercially-available dose calibrator with a
modified shielding liner. Xenon-133 was used as a first test, since a gas is
expected to provide the most homogeneous distribution within the balloon,
without concern for air bubbles that might be present in liquid sources. A
special shielding insert was constructed that would leave the balloon exposed,
while shielding the dose calibrator chamber from radiation in the catheter.
The dose calibrator was calibrated for 133Xe in the appropriate
geometries at NIST. The entire assembly was then tested by assaying balloon
catheters (3 mm diameter × 20 mm length or 3.5 mm diameter
× 20 mm length) for the amount of contained 133Xe.
GAFChromicTM film measurements were performed using balloon
catheters that had just been assayed for the amount of contained activity of
133Xe and the results compared to Monte Carlo calculations. Because
of the low energy of the 133Xe radiation, many source
inhomogeneities were evident, leading to large (~50 %) uncertainties.
Despite this, agreement with Monte Carlo calculations is good and should
improve with higher-energy beta-emitters. (B.E. Zimmerman,
M.P. Unterweger, C.G. Soares, and S.M. Seltzer, and
R.C. Chan [WHC])
- Approaches in Intravascular Brachytherapy. We developed a system of
entrapping 99mTc within liposomes of various charged (positive,
negative or neutral) membranes to be used as a possible, technetium-packaging
unit for intravascular irradiation (IVBT) after balloon angioplasty (using the
InfiltratorTM catheter system). After preliminary labeling of
liposomes showed the feasibility of this approach, labeling of liposomes with
99mTc at the Washington Hospital Center (WHC) was performed.
Labeled liposomes were injected into pig models and a time course was followed
over three hours for persistence of liposomes at artery sites (both with and
without injury). Fuji imaging of excised arteries (with free 99mTc
and 99mTc-labeled liposomes) showed that 99mTc-labeled
liposomes persisted longer at the application site than free 99mTc.
We have demonstrated that liposome encapsulation of 99mTc is highly
efficient at volumes up to 1 mL and that the liposomes are stable (do not
leak) under in vivo types of conditions (in presence of serum) although
stannous chloride is necessary for efficient encapsulation. Work has begun to
determine the feasibility of using other 99mTc-labelled compounds
(SestamibiTM, CeretecTM) in the same system.
(L.R. Karam, B.E. Zimmerman, with R. Waksman and R. Chan
[WHC])
- Determination of Calibration Factors for the Nondestructive Assay of
Pure Beta-emitting Brachytherapy Sources. Ionization-chamber calibration
factors have been determined for two, commercially-manufactured,
intravascular, brachytherapy sources: a TiNi-encapsulated 32P
source developed by Guidant Intravascular Intervention (Houston, TX), and a
stainless-steel encapsulated 90Sr-90Y developed by Bebig
Isopentechnic und Umweltdiagnostik GmbH (Berlin, Germany) in collaboration
with the Novoste Corp. (Norcross, GA). The calibration factor for the former
was derived from ionization current measurements with a Capintec CRC-12
("dose calibrator") which is the nuclear-medicine community's de facto
standard instrument, followed by very quantitative, destructive assays of the
32P content in the sources. Similarly, for the former sources, NIST
"4πγ chamber A" calibration factors were determined
for both bare ceramic and stainless-steel encapsulated source configurations.
(R. Collé)
- Development of Procedures for the Chemical Digestion and Radionuclidic
Assay of Encapsulated, Pure Beta-Emitting, Intravascular Brachytherapy
Sources. These methods have been applied to two very different types of
sources: a polymer-based, 32P-containing source with TiNi
encapsulation; and a ceramic-based, 90Sr-90Y source with
stainless-steel encapsulation. The internal compositions for both types of
sources had constituents that were chemically impervious. The assays involved
partial dissolutions (or extractions) followed by 4πβ liquid
scintillation (LS) spectrometry (with 3H-standard efficiency
tracing) of the resulting solutions. The procedures for the former included
provisions for accounting for all possible losses of 32P in the
digestion procedure (based on radiochemical tracing experiments), for any
unrecovered activity in the remaining source material, and for any residual
activity in the solution- and source-handling tools. The procedures for the
latter source consisted of extracting a fraction of the 90Sr
activity from the ceramic-like material for LS assay, and determining the
fraction of unextracted activity by before and after ionization current
measurements on the extracted source material. The uncertainties in the assays
were typically 2 % to 4 % for two standard uncertainty intervals. These
destructive assays were required for relating radiochromic-film measurements
of the absorbed dose spatial distributions for the sources to theoretic dose
modeling, and for establishing calibration factors for subsequent,
non-destructive, radionuclidic measurements on the sources.
(R. Collé)
- New "Primary" Standardizations of 226Ra and
222Rn. All extant activity standards and calibrations for
226Ra and 222Rn are based on comparative measurements
against an artifact radium mass standard (namely the international 1935
Hönigschmid standards and its derivatives). Efforts are currently
underway to remove this dependence and to develop and perform primary
standardizations for both 226Ra and the 222Rn subseries
based on 4παβ LS spectrometry with 3H-standard
efficiency tracing. We have successfully employed this new standardization
technique to 222Rn, such as for the calibration of the NIST
radon-in-water standard generator and for measurements of the
polyethylene-encapsulated 226Ra-solution emanation standards
(SRM 4968). The techniques are now being extended to the entire
226Ra decay series. Preliminary findings have demonstrated that the
method will be able to adequately resolve the 210Pb subseries
(i.e., from 226Ra and the 222Rn subseries) in even aged
radium solutions. The ability to perform such a resolution is significant in
that it will allow direct intercomparisons between the various 226Ra
standards issued by NIST over the past 50 years, irrespective of their
present degree of radioactive equilibrium between 226Ra and
210Pb. (R. Collé)
- Standardization of New Tritiated-Water Standards. The calibration
of the new, tritiated-water standard, SRM 4927F, by internal-gas counting
and liquid scintillation counting is in progress. This standard and the lower
activity SRM 4962F will be used extensively worldwide for ground-water
hydrology. The standards and their predecessors are and will be used as the
basis for a uniform measurement scale for environmental tritium studies. The
NIST tritiated-water standards are also used extensively in liquid
scintillation counting. A redetermination of the half-life of tritium is also
being done. An accurate value for the half-life is very important in extending
the useful lifetime of the tritium standards, i.e., the decay of the certified
value is not significantly effected by the uncertainty in the half-life value.
(L.L. Lucas and M.P. Unterweger)
- Titanium-44 Half-life. The 1157 keV gamma ray resulting from
the decay of 44Ti (and its 3.93 hour daughter 44Sc)
has been observed from the supernova remnant, Cassiopeia A. The flux of
this gamma ray and the 44Ti half-life would yield a production
abundance of the isotope at the time of the explosion. This is an important
"data point" for the test of theoretical models of supernova events.
Historically, the measured 44Ti half-life values have ranged from
39 years to 66 years, much too uncertain to be useful.
The Radioactivity Group has had 44Ti sources since 1978. The
emission rate of the 1157 keV gamma ray from these sources has been
measured from time to time with the Group's carefully calibrated, gamma-ray
detector systems. These emission rate measurements were carefully assembled
and documented. Over the last year, additional emission rate measurements
have been made and a half-life (60.7 years, with an estimated standard
error of 1.2 years) determined. (F.J. Schima, F.E. Wietfeldt, and
B.M. Coursey)
Most Recent Technical Activities
|
Archive of Technical Activities
|